Call Toll Free (888) 310-7322

ESA609 Electrical Safety Analyzer

The ESA609 Electrical Safety Analyzer is a rugged, portable and easy-to-use analyzer designed for general electrical safety testing. Engineered for on-the-go technicians, the ESA609 requires no prior training, has a rubberized case that allows it to sustain the rigor of transportation, and helps prevent damage when accidentally dropped.

Description

Its functional strap and featherweight design make it one of the most portable electrical safety analyzers in its class. Heavy-duty switches allow you to effortlessly change polarity and configuration of the neutral connection between open and closed, while push-button operation ensures fast transition between tests for complete basic testing in minutes. The ESA609 integrates all functions needed to test medical devices when patient lead testing is not required, including: line (mains) voltage, ground wire (protective earth) resistance, equipment current, leakage current and point-to point tests. Versatile to global electrical safety standards of choice, the ESA609 tests to ANSI/AAMI ES1, NFPA-99, and parts of IEC62353 and IEC60601-1.

 

Key Features

  • IEC Standards compliance: ANSI/AAMI ES1, NFPA-99, and parts of IEC62353 and IEC60601-1
  • Portable: Featherweight design, functional strap, and tilt stand make it easy for transportation and operation on-the-go (onsite or offsite)
  • User-friendly: Quick push-button operation for rapid testing
  • Rugged: Rubberized case and Ingress Protection rating of IP30 help prevent damage when dropped
  • Global use: the ESA609 will operate at 120V and 230V
  • Test current consumption up to 20A for a diverse set of medical devices
  • All parameters needed for basic electrical safety testing: Line (Mains) voltage, Ground Wire (or Protective Earth) Resistance, Equipment current, Ground Wire (Earth) leakage, Chassis (Enclosure) leakage, Direct equipment leakage, and Point to point leakage and resistance
  • Rigorously tested for safety and reliability, with CE, CSA and Australia RCM in addition to Fluke quality
  • Two-year extended warranty (no-cost, available after first-year calibration at any authorized Fluke Biomedical Service Center)
  • Global support network delivering prompt service and peace of mind to Fluke Biomedical customers worldwide

Specifications

General Specifications

Display: LCD display
Modes of Operation: Manual
General: Answer

Temperature

Storage: -20 °C to 60 °C (-4 °F to 140 °F)
Humidity: 10 % to 90 % non-condensing

Altitude

120 V ac mains supply voltage: 5000 m
230 V ac mains supply voltage: 2000 m

Power Ratings

115 Volt power outlet: 90 to 132 V ac rms, 47 to 63 Hz, 20 A maximum
230 Volt power outlet: 180 to 264 V ac rms, 47 to 63 Hz, 16 A maximum
Power input: 115V 20A – 2.6kVA and 230V at 16A – 4.2kVA

Physical case

Weight: 0.7 kg (1.5 lb)
Dimensions (WxDxH): 22.9 cm x 17.8 cm x 6.4 cm (9 in x 7 in x 2.5 in)

Environmental

Agency Approvals: CE, CSA, Australia RCM
Safety: IEC 61010-1:Overvoltage Category II, Measurement 300V CAT II, Pollution Degree 2
Electromagnetic Environment: IEC 61326-1:Portable
Safety: IEC CISPR 11:Group 1, Class A.

  • Group 1 have intentionally generated and/or use conductively coupled radio-frequency energy which is necessary for the internal functioning of the equipment itself.
  • Class A equipment is suitable for use in nondomestic locations and/or directly connected to a low-voltage power supply network

Warranty

Warranty: Two-year extended warranty (no-cost, available after first-year calibration at any authorized Fluke Biomedical Service Center, otherwise standard one year warranty applies)